Metrohm 797 VA Computrace Manual do Utilizador Página 254

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 253
8 How to ...?
797 VA Computrace – Software
242
Es wird für beide Dosino ein "Prep-Zyklus" empfohlen.
On the Automation tab, select the 838 Advanced Sample Processor.
The default settings can be adopted (click button <Default>).
Method Parameters for the 797
Two steps are required for Suppressor determination. One to re-
cord the response curve with Calibration technique "RC Record re-
sponse curve", another for the actual sample determination with
Calibration technique "RC Sample with response curve". Set CVS or
CPVS for Mode for both methods.
Recording of the Response Curve::
Choose "RC Record response curve" for
Calibration.
Settings in the Dosinos window:
After the addition of Electrolyte solution, Suppressor standard
solution is added via Dosino 3. That should be specified on the
Substances tab of the EDIT WORKING METHOD PARAMETERS
window (here the window for CVS):
Vista de página 253
1 2 ... 249 250 251 252 253 254 255 256 257 258 259 ... 294 295

Comentários a estes Manuais

Sem comentários