8 How to ...?
797 VA Computrace – Software
236
Activate both options on the General tab of the GENERAL SET-
TINGS
window, so that a recorded calibration curve replaces the
old one automatically.
Settings on the Dosinos tab of the GENERAL SETTINGS window in
the 797 Software:
It is recommended to do for Dosino1 one "Prep cycle" and for Dos-
ino3 two "Prep cycles".
On the Automation tab, select the 838 Advanced Sample Processor.
The default settings can be adopted (click button <Default>).
Method Parameters for the 797
Two steps are required for Suppressor determination. One to re-
cord the calibration curve with Calibration technique "DT Record
calibration curve", another for the actual sample determination
with Calibration technique "DT Suppressors with calibration curve".
Set CVS or CPVS for Mode for both methods.
Recording of the calibration curve:
Choose "DT Record calibration curve" for Calibration.
Settings in the Dosinos window: