6.26 838 Advanced Sample Processor
IC Net 2.3
237
Alias Define the alias name which appears in the
rest of the driver.
Default: Stirrer MSB1…3, Tower Stirrer
Rate
Define stirring rate.
Entry range: 1, … , 15 default 3
<Set> Send parameters to the devices.
Membrane pumps
EP1 Configure membrane pumps.
Alias Define the alias name which appears in the
rest of the driver.
Default: Membrane pump 1…2
838 Autoinit
To open the 838 Advanced Sample Processor AUTOINIT window,
click with the right mouse button on the 838 icon, and choose
autoinit.
If the Autoinit option of the 838 Advanced Sample Processor is
activated, all instrument parameters are set to the default value
each time the instrument is switched on.
Deactivate the option Autoinit before you disconnect the 838
Advanced Sample Processor to control it with a keyboard.
To activate Autoinit: Check the box Autoinit and click on <Set>
(while instrument is connected).
To deactivate Autoinit: Uncheck the box Autoinit and click on
<Set> (while instrument is connected).
Exit the window with the <OK> button, to make sure that chan-
ges made are saved in the method.