Metrohm 827 pH lab Manual do Utilizador Página 50

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 82
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 49
7.3 Setup
42 826/827 pH meter, Instructions for Use
7.3 Setup
Certain settings of the pH meter can be found in the setup menu. This
menu cannot be accessed from the normal instrument display, as alte-
rations to these settings have a great influence on the functions of the
pH meter.
The setup menu is called up by pressing down and holding down the
<CAL> key while switching on the pH meter.
setup
lock
instr.number
instr.type
serial type
power type
Attention!
Be very careful when making alterations in this menu! Alterations have
a very large influence on the functions of the instrument. Accidental
alterations can be rejected by carrying out a deliberate RAM initializa-
tion (see Section 8.3).
7.3.1 Locking
Setup
lock
: all funct.
<CAL>
With all funct. you can disable all functions as well as the <CAL>
key, e. g. to prevent unwanted alterations to the instrument settings.
The measuring mode can be changed nevertheless. The <CAL> key
can be locked separately.
You can activate such a disablement by setting the corresponding pa-
rameter to 'on'. This does not affect access to the Setup menu in any
way.
Please note that all disablements will be invalidated if a RAM-Init is car-
ried out (see Section 8.3).
Vista de página 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 81 82

Comentários a estes Manuais

Sem comentários