7 Methods
IC Net 2.2
178
parameter is used to detect an underflow
condition.
Zero A signal level on a baseline (in ADC con-
version units). This parameter is a result of
the ADC calibration.
Maximum Maximum value of the linear range of the
AD interface (in ADC conversion units). This
parameter is used to detect overflow condi-
tion.
Range An input signal value in maximum in user-
defined units.
Range = (Maximum - Zero) Coef
Coefficient ADC sensitivity coefficient (weight of ADC
bit in user-defined units).
Noise Estimated baseline noise value of the
channel in ADC conversion units (bits).
Shift Shift of the current channel in time. This pa-
rameter is useful in the case of multi-
channel chromatograms only.
(multi)Channel ratio Spectral ratios. The option is valid for multi-
channel chromatograms only.