6.11 766 IC Sample Processor
IC Net 2.0
91
Num Row number (read-only).
Task Program instruction (see below).
Parameter Parameter for program instruction (see
below).
Enabled Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program tab:
Instruction Parameter entry Meaning
Move sample, sample+1,
spec.#, 1...n
Move rack to place the specified
vessel at the needle.
Lift shift, work, rinse,
special,
0, 1...125 mm
Raise or lower the lift to the specified
position.
Pump on, off, 1...999 s Switch on/off peristaltic pump at
the tower.
Scan "scan command" Scan remote interface until the
specified signal is received. In ad-
dition to the normal scan com-
mands (details see 766 Instruc-
tion for Use) special commands
defined on the Configuration tab
can be used.