5.2 Maintenance
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
IC Equipment Thin-Layer cell
1
Unscrew the pressure screw (1-4) on the base part of the cell and
remove it.
2
Take out the working electrode.
3
Insert the new working electrode. The working electrode is formed in
such a way that it can be inserted into the measuring cell in only one
position.
4
Slide the pressure screw over the working electrode and screw it
tightly.
5.2.2 Replacing the reference electrode
Note
These instructions apply analogously to the Ag/AgCl reference elec-
trode.
Inserting the reference electrode
Prerequisites:
The measuring cell is switched off.
The measuring cell has been removed from the detector.
The electrode cables are unplugged.
No tool is required for replacing the reference electrode.
1
Slide the supplied sealing ring over the new reference electrode.