3.1 Preparing the ultrafiltration cell
■■■■■■■■■■■■■■■■■■■■■■
8
■■■■■■■■
IC equipment: Inline-Ultrafiltration
3
Disassembling the ultrafiltration cell
■ Remove the upper part of the ultrafiltration cell.
■ Remove the sealing ring.
4
Cleaning the ultrafiltration cell
CAUTION
Damage to the ultrafiltration cell
Organic solvents (e.g. acetone) corrode and damage the ultrafiltra-
tion cell material (PMMA).
■ Use only ultrapure water or a water-ethanol mixture (70:30) for
cleaning the ultrafiltration cell.
■ Thoroughly rinse off the sealing ring (2-3), the upper chamber
(2-2) and the lower chamber (2-4) of the ultrafiltration cell with
ultrapure water.
■ Dry all parts with a lint-free cloth.