2.6 High-pressure pump
792 Basic IC
25
2.6.3 Connecting the eluent bottle
To connect the eluent bottle with the high-pressure pump, insert the
supplied 6.1834.010 aspirating tubing (i.d. = 1.5 mm, e.d. = 2.5 mm,
length = 2.5 m) into one of the openings
5
55
5
or
7
77
7
in the interior of the 792
Basic IC. Pull the aspirating tubing sufficiently far into the interior of the
792 Basic IC and push at least 5 mm of it onto aspirating capillary
32
3232
32
(see
Fig. 3
) of the high-pressure pump (it may be necessary to use em-
ery paper). Screw on the 6.2821.090 aspirating filter at the other end of
the aspirating tubing and insert it into the eluent bottle.
Only degassed (with N
2
, He or vacuum) and microfiltered (0.45
P
m
filter) eluents should be used!
Care must be taken that the eluent used is freely miscible with any
solvent remaining in the pump head (the pump head leaves the
factory filled with either isopropanol or methanol/water). If this is not
the case then the pump must first be rinsed with a solvent which is
miscible with both the previous eluent and the following eluent (e.g.
acetone).
2.6.4 Deaerating the pump and rinsing the pulsation dampener
The first time that it is started up the high-pressure pump must be
deaerated. Proceed as follows:
1 Prepare for deaeration
x
Open the rotary knob on purge valve
31
3131
31
by approx. ½ turn in
the counterclockwise direction (see
Fig. 3
).
x
Mount a PEEK compression fitting
46
4646
46
(see
section 2.3.4
) to
the end of capillary
30
3030
30
and screw coupling
24
2424
24
(6.2744.120) to
the compression fitting
46
4646
46
.
x
Push 6.2816.020 Syringe (without needle) into the connection
on coupling
24
2424
24
until the stop is reached.
2 Open system
x
Start the
«
792 Basic IC
»
PC program, if it has not been al-
ready been started (see
section 2.5.3
).
x
Select
File / Open / System
in the main window and open the
folder
Default
. Select the system file
Default.smt
and click on
<Open>
. The corresponding system window is opened: