CH-9101 Herisau/Switzerland E-Mail [email protected] Internet www.metrohm.com 775 Dosimat Instructions for Use
2.2. Diagnosis 775 Dosimat 6 2.2.2 Summary 2.2.3 RAM-test ...6 2.2.4 D
2.2 Diagnosis 775 Dosimat 7 2.2.5 Diagnosis of spindle drive 1.Insert exchange unit and put the burette tip into a collecting receptacle. 2.Connect
2.3. RAM-initialisation 775 Dosimat 8 2.3 RAM-initialisation In rare cases, it is possible that major interference signals such as line spikes a
2.4 Releasing a locked spindle 775 Dosimat 9 2.4 Releasing a locked spindle with inserted Exchange Unit The burette drive may very occasionally ja
3.1. Technical specifications 775 Dosimat 10 3 Appendix 3.1 Technical specifications Exchange units 1, 5, 10, 20, 50 mL burette cylinder volum
3.1 Technical specifications 775 Dosimat 11 Power supply Voltage 100, 117, 220, 230/240 V ± 10% (adjustable) Frequency 50... 60 Hz Consumpt
3.2. Warranty and certificates 775 Dosimat 12 3.2 Warranty and certificates 3.2.1 Warranty The warranty regarding our products is limited to re
3.2 Warranty and certificates 775 Dosimat 13 3.2.2 Certificate of Conformity and System Validation This is to certify the conformity to the stand
3.2. Warranty and certificates 775 Dosimat 14 Ionenanalytik • Analyse des ions • Ion analysis • Análisis iónico 775 Dosimat EU- Declarati
3.3 Scope of delivery and ordering designations 775 Dosimat 15 3.3 Scope of delivery and ordering designations Dosimat 775 ...
Index 775 Dosimat 16 Index Keys are marked with < >, display texts are in bold characters. A Accessories...
Content 775 Dosimat Content 1 Overview ...
Overview 775 Dosimat 1 Explanation of symbols: < > means "key", e.g. <GO> means key "GO" DOS...0.000
Overview 775 Dosimat 2 1 Overview Front view of instrument: 1 Exchange unit Choose the volume of the Exchange unit in such a way that a v
Overview 775 Dosimat 3 Rear view of instrument: 6 Identification plate Indication of model, series and serial number. 7 Indication of powe
2.1. Special messages and error messages 775 Dosimat 4 2 Error messages, troubleshooting 2.1 Special messages and error messages error 1
2.2 Diagnosis 775 Dosimat 5 2.2 Diagnosis 2.2.1 General The 775 Dosimat is a very precise and dependable feeding instrument. Thanks to its rugged c